| It is a fact is that environmental air enters in the Genie®'s balloon. How can I be sure the medication will not get contaminated? |
| Genie®'s balloon is made from silicone, which is permeable to gases. The air entering the balloon slowly passes through the silicone and empties out the balloon. During evaluations performed on Genie®, it was shown that even the Phi-X174 bacteriophage, one of the smallest known viruses (smaller than drug vapor particles), cannot cross the barrier of the silicone balloon. |
| |
| Do I need any special syringe to use the Oncology Preparation and Delivery System? |
| Any male luer syringe, either luer lock or luer slip type, which complies with the ISO standard, is suitable for use with the Oncology Preparation and Delivery System? |
| |
| We have a Spiros™ attached to the proximal end of the IV line. Is it mandatory to connect Spiros™ to a CLAVE® connector? |
| No. Even though connection to a CLAVE®connector is highly recommended, it is not the only connector that will activate Spiros™. Any female Luer connection may be used to activate Spiros™. |