Safe handling made simple
For the Safe Preparation of Hazardous Drugs |
|
 |
|
| |
| TECHNICAL SPECIFICATIONS: SPIROS ™ |
| |
| Application |
| The Spiros™ connector is a luer activated closed system, to be used in the preparation, transport and administration of hazardous drugs. |
| |
| Operation |
| The female Luer connector on Spiros™ attaches to a syringe or IV line, becoming permanently fixed. When the male Luer connection on Spiros™ is connected to a needle-free system, the internal channel opens allowing fluid to flow freely through it. Upon disconnection, Spiros™ generates a negative internal pressure which absorbs the liquid remaining between the connectors, keeping it within the system. Spiros™prevents distal dripping and can only be activated by a needle-free system with a standard ISO female Luer. |
| |
| Connector Compatibility |
| Compatible with standard Luer connectors in conformity with the ANSI (American National Standards Institute) for Luer connectors under ISO 594 and ANSI / HIMA MD 70.1-83 guidelines. |
| |
| Materials |
| |
Body: |
Polycarbonate |
| |
Seals: |
Silicone |
| |
Central Spike: |
Polycarbonate |
| |
All materials comply with USP Class IV guidelines compliant and are latex or DEHP free. |
| Back Pressure: |
<20 psig (<137.9kPa) |
| Priming volume: |
0.1 ml |
| Residual Volume: |
0.1 ml |
| Flowrate: |
126 ml/min |
| Leakage during use: |
None |
| Extended use: |
No leaks after 10 activations |
| Negative Displacement: |
0.01-0.03 ml |
| Compatibility: |
Compatible with alcohol or lipid soluble medications. |
|
| |
| |
|
|
|
| |
| |