| Positive Displacement Connector |
| |
| Preventing catheter occlusion |
|
 |
|
| |
| TECHNICAL SPECIFICATIONS |
| |
| Application |
| The CLC2000® Needle-Free Connector is an injection port that is activated without the need of needles for IV applications. It can be used for aspiration as well as for injection of fluids. |
| |
| Flow Path |
| The CLC2000® design ensures internal flow, in which the liquids that pass through the connector never come into contact with any component exposed to the environment, creating a closed system, mechanically as well as microbiologically. |
| |
| Compatible connectors |
| Standard Luer connectors compatible with the American National Standards for Luer connectors, under the directives ISO 594 and ANSI / HIMA MD 70.1-83. |
| |
| Materials |
| Body: Valox® (Polycarbonate & Fiberglass) |
| Seals: Silicone |
| Piston: Polycarbonate |
| Activation Spring: Stainless Steel |
| All materials comply with USP Class IV directives and do not contain latex or DEHP. |
| |
| Back Pressure: |
60 psig (413.7kPa) |
| Priming volume: |
0.06 ml |
| Residual Volume: |
0.0 ml |
| Flow: |
180 ml/min |
| Positive Displacement: |
0.035 ml |
| Leakage during use: |
None |
| Extended use: |
No leaks after 1,000 activations in 10 days. |
| Microbial Barrier: |
Repeated accesses according to rule SE00-0081T for 7 days. |
| Compatibility: |
Compatible with Cytostatics, Lipids (TPN) and blood or blood products. |
|
| |
|
|
|
| |
| |