| TECHNICAL SPECIFICATIONS |
| |
| Application |
| The CLAVE® Needle Free Connector is an injection port activated without the need of needles for IV applications. It can be used in the aspiration as well as the infusion of fluids on demand. |
| |
| Flow Path |
| The CLAVE® design ensures an internal flow path, where liquids that pass through the connector never come into contact with any component exposed to the environment, creating a mechanically and microbiologically closed system. |
| |
| Connectors |
| Standard luer connectors compatible with the American National Standards for Luer connectors, under ISO 594 and ANSI / HIMA MD 70.1-83 standards. |
| |
| Materials |
| Body: Valox® (Polycarbonate + Fiberglass). |
| Seal: Silicone. |
| Central Spike: Polycarbonate. |
| All materials comply with USP class IV standards and do not contain latex or DEHP. |
| |
| Backpressure: |
60 psig (413.7kPa) |
| Maximum operating pressure: |
3000 mmHg |
| Priming volume: |
0.06 ml |
| Residual Volume: |
0.02 ml |
| Flow: |
185 ml/min |
| Leakage during use: |
None |
| Extended use: |
No leakage after 144 activations in 144 hours. |
| Microbial Barrier: |
Repeated accesses using P. aeruginosa, approved for 24 activations per day for 7 days. With S. epidermidis, approved for 8 activations per day for 7 days. |
| Compatibility: |
Compatible with cytostatics, lipids (TPN) and blood or blood products. |
|
| |
| |
| |
| |
|
|
|
| |
| |