Needle-Free,
Free of Complications |
|
 |
|
| |
| TECHNICAL SPECIFICATIONS |
| |
| Application |
| The Needle-Free Connector Flux® is an injection port that is activated without the need of needles for IV applications. It can be used for aspiration as well as for the infusion of fluids on demand. |
| |
| Connectors |
| Standard Luer connector. It provides a tight seal between the Luer tip of the syringe or IV system and the connector. All connectors are compatible with the American National Standards for Luer connectors under the directives ISO 594 and ANSI / HIMA MD 70.1-83. |
| |
| Materials |
| Body: Polycarbonate. |
| Seal: Silicone. |
| All components are sealed by ultrasound. |
| All materials comply with the USP Class IV directives and do not contain latex or DEHP. |
| |
| Back Pressure: |
40 psig (275.8kPa) |
| Activation Force: |
1.9 lb (0.86 kg) |
| Priming volume: |
0.09 ml |
| Residual Volume: |
0.11 ml |
| Flow: |
360 ml/min @ 1 psig (6.9 kPa)
600 ml/min @ 3 psig (20.7 kPa)
750 ml/min @ 5 psig (34.5 kPa) |
| Leakage during use: |
None at 1, 5 and 30 psig (6.9, 34.5 and 206.8 kPa) |
| Multiple Activations: |
No leaks after 100 activations at 36 psig (248.2 kPa). |
| Extended use: |
No leaks after 24 activations at 36 psig (248.2 kPa). |
| Microbial Barrier: |
Repeated accesses using 70% isopropyl ilcohol as a disinfectant and S. aureus as the offending organism. It is approved for 12 activations in 3 days and 140 activations within 7 days. |
| Biocompatibility: |
Demonstrated biocompatible, according to the evaluation procedures of standard ISO10993-1. |
|
| |
| |
| |
| |
|
|
|
| |
| |