Neutral Displacement Connector |
|
 |
|
| |
| TECHNICAL SPECIFICATIONS |
| |
| Application |
| The MicroCLAVE® Needle-Free Connector is an injection port activated without needles for IV applications. It can be used for aspiration as well as for the infusion of fluids on demand. |
| |
| Flow Path |
| The MicroCLAVE® design ensures an internal flow, in which the liquids that pass through the connector never come into contact with any component exposed to the environment, creating a closed system, mechanically as well as microbiologically. |
| |
| Displacement |
| When disconnecting the syringe or IV system from the MicroCLAVE®, the connector ensures that blood does not enter into the catheter, thus avoiding the formation of clots and having to change catheters. |
| |
| Connectors |
| Standard Luer connectors compatible with the American National Standards for Luer connectors, under ISO 594 and ANSI / HIMA MD 70.1-83 standards. |
| |
| Materials |
| Body: Valox® (Polycarbonate & Fiberglass) |
| Seal: Silicone |
| Central Spike: Polycarbonate |
| All materials comply with USP Class IV directives and do not contain latex or DEHP. |
| |
| Backpressure: |
75 psig (517.1kPa) |
| Priming volume |
0.04 ml |
| Residual Volume: |
0.04 ml |
| Flow: |
165 ml/min |
| Leakage during use: |
None |
| Extended use: |
No leakage after 144 activations in 144 hours. |
| Microbial Barrier: |
Repeated accesses using P. aeruginosa, approved for 24 activations per day for 7 days. With S. epidermidis, approved for 8 activations per day for 7 days. |
|
| |
| |
| |
| |
|
|
|
| |
| |
|
| |