TECHNICAL SPECIFICATIONS
 
Application
TEGO® Closed System for Hemodialysis is a connector activated by the Luer on the hemodialysis line that keeps the working circuit totally closed off to the exterior before, during, and after the procedure.
 
Channel Flow
The design of TEGO® ensures a high internal flow, in which the blood that passes through the connector never comes into contact with any components exposed to the environment, creating a meachincally and microgiologically closed system.
 
Connectors
Standard Luer connectors compatible with the American National Standards for Luer connectors under the directives ISO 594 and ANSI / HIMA MD 70.1-83.
 
Materials
Body: Valox® (Polycarbonate & Fiberglass)
Seal: Silicone
Cover: Silicone
All materials comply with USP Class IV directives and do not contain latex or DEHP.
 
Backpressure > 15 psig (> 103.4 kPa)
Priming volume 0.06ml
Flow 600 ml/min
Displacement Neutral
Leakage during use None
Extended use Without leaks after 4 activations per day during 7 days.
Microbial Barrier Approved according to protocol 04-256T for 7 days.
Compatibility: Compatible with alcohol, Lipids (TPN) and blood or blood products.